The 220DA Baratron® is temperature controlled to 45°C for use in demanding, harsh environments. The Type 220DA includes the signal conditioner and power supply, and is enclosed in a rugged, NEMA1 approved housing. It is available in ranges from 1 to 25,000 mmHg/Torr (500 psia). Accuracy is ±0.15% of Reading.
- Temperature controlled to 45°C for 0.15% of reading accuracy
- Simultaneous 0-10 VDC and 4-20 mA outputs proportional to absolute pressure
- NEMA 1 enclosure for sensor protection in harsh industrial environments
- Usable resolution of 1 part in 10,000 covers a wide range of pressures See All Features
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | 220DA-00001A2BAbsolute Pressure Transducer, 45°C, 1 Torr, 0.15%, 0.5 in. Tube, NEMA 1 | |||||
![]() | 220DA-00001B2BAbsolute Pressure Transducer, 45°C, 1 Torr, 0.15%, 8 VCR-F, NEMA 1 | |||||
![]() | 220DA-00001D2BAbsolute Pressure Transducer, 45°C, 1 Torr, 0.15%, NW16 ISO-KF, NEMA 1 | |||||
![]() | 220DA-00010A2BAbsolute Pressure Transducer, 45°C, 10 Torr, 0.15%, 0.5 in. Tube, NEMA 1 | |||||
![]() | 220DA-00010B2BAbsolute Pressure Transducer, 45°C, 10 Torr, 0.15%, 8 VCR-F, NEMA 1 | |||||
![]() | 220DA-00010D2BAbsolute Pressure Transducer, 45°C, 10 Torr, 0.15%, NW16 ISO-KF, NEMA 1 | |||||
![]() | 220DA-00100A2BAbsolute Pressure Transducer, 45°C, 100 Torr, 0.15%, 0.5 in. Tube, NEMA 1 | |||||
![]() | 220DA-00100B2BAbsolute Pressure Transducer, 45°C, 100 Torr, 0.15%, 8 VCR-F, NEMA 1 | |||||
![]() | 220DA-00100D2BAbsolute Pressure Transducer, 45°C, 100 Torr, 0.15%, NW16 ISO-KF, NEMA 1 | |||||
![]() | 220DA-01000A2BAbsolute Pressure Transducer, 45°C, 1000 Torr, 0.15%, 0.5 in. Tube, NEMA 1 |
Configuration Options
The following options are available for 220DA Baratron® Absolute Pressure Transducers
Ordering Code Example: 220DA00100A2B
| Configuration Option | Option Code |
|---|---|
| 220DA Absolute Capacitance Manometer | 220DA |
Range Full Scale |
|
| 1 Torr | 00001 |
| 10 Torr | 00010 |
| 20 Torr | 00020 |
| 100 Torr | 00100 |
| 1000 Torr | 01000 |
| 5000 Torr | 05000 |
| 10000 Torr | 10000 |
| 15000 Torr | 15000 |
| 20000 Torr | 20000 |
| 25000 Torr | 25000 |
Fittings |
|
| 0.5 in. OD tube | A |
| 8 VCR female | B |
| 1.33 in. OD CF, rotatable | C |
| NW16 ISO-KF | D |
Specifications
- TypePressure Transducer
- Sensor TypeIndustrial Capacitance Manometer
- Measurement TypeAbsolute
- Full Scale Pressure Ranges1, 10, 20, 100, 1000, 5000, 10000, 15000, 20000, 25000 Torr
- Sensor TemperatureControlled to 45°C
- Accuracy0.15% of Reading
- Analog Output0 to 10 VDC into > 10K Ω load and 4 to 20 mA into < 500 Ω load
- Volume7.0 cc
- Overpressure Limit35 psia
- Power Requirements115/230 VAC, 50/60 Hz
- Enclosure RatingNEMA 1
Features
Baratron® Capacitance Manometer Technology
Capacitance manometers are electro-mechanical gauges that can measure both pressure and vacuum. The capacitance gauge translates a pressure-modulated movement in a thin diaphragm into an electrical signal proportional to the pressure. The pressure sensor is the thin diaphragm that is exposed to the pressure or vacuum being measured via the inlet tube. An electrode is mounted in the reference cavity behind the diaphragm. Pressure differences between the process and the reference cavity deflect the diaphragm slightly, changing the distance between it and the electrode. Variations in this distance produce variations in the capacitance between the diaphragm and the electrode creating an electrical signal that is proportional to the pressure change. Since differences in the capacitance signal are produced by physical changes within the manometer and not by changes in the gas properties, pressure measurements by the capacitance manometer are independent of the composition of the gas being measured.
Internally Heated to 45°C
These capacitance manometers are temperature controlled to 45°C for improved accuracy. Unheated versions are exposed to ambient temperature variations which can degrade the sensor accuracy. These devices have the sensor enclosed in a volume that is maintained at a constant temperature above ambient. This solution improves the manometer’s accuracy and repeatability and lowers instrument drift by reducing or eliminating process contamination within the manometer. Heated manometers are recommended for applications that require maximum accuracy and repeatability, operate above ambient temperature and for those processes that employ hot gases.
Absolute Pressure Measurement
These Baratron® pressure transducers are referenced to vacuum for absolute pressure measurement. Applications include: vacuum furnaces, freeze-drying of fruits and vegetables, gas lasers, automotive component testing, bottle coatings, and vacuum distillation.
All-Inconel® Wetted Surfaces
These pressure transducers feature all Inconel® construction of the pressure sensor allowing it to operate without damage in virtually any chemical environment, including exposure to halogens, deionized water and steam, and ozone.
0-10 VDC and 4-20 mA Pressure Outputs
These Baratron® capacitance manometers feature two simultaneous 0 to 10 VDC and 4 to 20 mA analog pressure outputs to interface with common process control systems. The 0-10 VDC signal allows it to be used as a pressure transducer, while the 4-20 mA signal supports use as a pressure transmitter.
NEMA 1 Rated Enclosure
All pressure sensor components and electronics are housed in a rugged, NEMA 1 rated enclosure for protection when used in harsh industrial environments.
Accessories
Compatible Cables
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | RCB112S-10-10 | |||||
![]() | RCB254S-1-10 | |||||
![]() | RCB474S-1-10 | |||||
![]() | RCB220S-4-10 |
Power Supplies and Displays
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | PR4000BF2V3 | |||||
![]() | PR4000BF5V3 | |||||
![]() | PR4000BF2V2 | |||||
![]() | PR4000BF5V2 | |||||
![]() | PR4000BS2V2 | |||||
![]() | PR4000BS2V3 | |||||
![]() | PDR2000AO |
Resources
Literature
220DA Baratron® Heated Capacitance Manometer Datasheet (202.4 kB, PDF)
Manuals
220D Baratron® Capacitance Manometer Manual(562.3 kB, PDF) 220 Capacitance Manometer Pin-outs(23.2 kB, PDF)