Compare Description Drawings, CAD & Specs Full Scale Range Fitting Type Electrical Connector Avail. Price
AA01A Baratron® 1-1000 Torr Unheated Absolute Capacitance ManometersAA01A01TGAS3B000001 TorrNW16 ISO-KF15-pin D-sub male with screw locks
$1,338
AA01A Baratron® 1-1000 Torr Unheated Absolute Capacitance ManometersAA01A11TBAS3B0000010 Torr0.5 in. OD tube15-pin D-sub male with screw locks
$1,242
AA01A Baratron® 1-1000 Torr Unheated Absolute Capacitance ManometersAA01A12TGCS3B00000100 TorrNW25 ISO-KF15-pin D-sub male with screw locks
$1,338
AA01A Baratron® 1-1000 Torr Unheated Absolute Capacitance ManometersAA01A13TGAS3B000001000 TorrNW16 ISO-KF15-pin D-sub male with screw locks
$1,338

Configuration Options

The following options are available for the AA01A Baratron® Absolute Capacitance Manometer.
Ordering Code Example: AA01A11TBAS3B00000

Configuration Option Option Code
AA01A Unheated Absolute Capacitance Manometer AA01A
Range Full Scale
1 01
2 02
10 11
20 21
100 12
500 52
1000 13
Engineering Units
Torr/mmHg T
mbar M
kPa K
Fittings (compatible with)
0.5 in. (12.7 mm) OD tube BA
4 VCR female (V sensor only) CD
8 VCR female CE
8 VCR male CF
NW16 ISO-KF GA
NW25 ISO-KF GC
1.33'' (33.8 mm) Conflat, rotatable HA
2.75'' OD Conflat, rotatable HC
8 VCO female DA
Sensor
Standard S
Short Standard (GC or HA fitting only) T
¼'' dia. tube (CD fitting only) V
Analog Output
0 to 10 VDC 3
Electrical Connector
15 POS D-subminiature-M with screw locks B
15 POS D-subminiature-M with slide lock posts P
Reserved
Reserved 00
Calibration Orientation
Ranges = 1 Torr 0
Ranges = 1 Torr, horizontal H
Ranges = 1 Torr, vertical V
Accuracy
Standard 0
Options
None 0
15 pin D terminal block adapter U
Description
AA01A Baratron® 1-1000 Torr Unheated Absolute Capacitance Manometers

AA01A

Absolute Capacitance Manometer, 1-1000 Torr, 0-10 VDC, DB15

Specifications

  • Type
    Pressure Transducer
  • Measurement Type
    Absolute
  • Full Scale Pressure Ranges
    1, 2, 10, 20, 100, 500, 1000 Torr and metric equivalents
  • Fitting Type
    1/2 inch OD tube standard, other types optional
  • Analog Output
    0 - 10 VDC into >10 kΩ load
  • Electrical Connector
    15-pin D-subminiature standard, terminal block adapter optional
  • Accuracy
    0.25% for 1 to 1000 Torr (0.15% optional)
  • Resolution
    0.002% of Full Scale
  • Zero Temperature Coefficient
    0.005% Full Scale/°C for 10-1000 Torr ranges
    0.010% Full Scale/°C for 2 Torr range
    0.015% Full Scale/°C for 1 Torr range
  • Span Temperature Coefficient
    0.04% of Reading/°C
  • Operating Temperature
    0 to 50°C
  • Overpressure Limit
    45 psia (310 kPa)
  • Volume
    1/2 inch Tube: 6.29 cm3
    8 VCR: 8.605 cm3
  • Materials Exposed to Process Gases
    Inconel and Incoloy nickel alloys
  • Power Requirements
    15 to 30 VDC @ 35 mA
  • Compliance
    CE

Features

Baratron® Capacitance Manometer Technology

Gas independent pressure measurement with a Baratron® capacitance manometer. Schematic shows the internal components and functional zones.

Capacitance manometers are electro-mechanical gauges that can measure both pressure and vacuum. The capacitance gauge translates a pressure-modulated movement in a thin diaphragm into an electrical signal proportional to the pressure. The pressure sensor is the thin diaphragm that is exposed to the pressure or vacuum being measured via the inlet tube. An electrode is mounted in the reference cavity behind the diaphragm. Pressure differences between the process and the reference cavity deflect the diaphragm slightly, changing the distance between it and the electrode. Variations in this distance produce variations in the capacitance between the diaphragm and the electrode creating an electrical signal that is proportional to the pressure change. Since differences in the capacitance signal are produced by physical changes within the manometer and not by changes in the gas properties, pressure measurements by the capacitance manometer are independent of the composition of the gas being measured.

Percent of Reading Accuracy

Accuracy is specified as a percent of Reading, not Full Scale, as seen in some of the lower performance devices. Percent of Reading accuracy provides you with an even more accurate output signal in the lower scale of the pressure range, where it is needed most.

Absolute Pressure Measurement

These Baratron® pressure transducers are referenced to vacuum for absolute pressure measurement. Applications include: vacuum furnaces, freeze-drying of fruits and vegetables, gas lasers, automotive component testing, bottle coatings, and vacuum distillation.

Inconel® and Incoloy® Construction

These pressure transducers feature Inconel® and Incoloy® nickel alloy construction of the pressure sensor allowing it to operate without damage in virtually any chemical environment, including halogens, deionized water and steam, and ozone. The sensor is fully welded with no intermediate brazing or joining materials.

Accessories

Power Supplies and Displays

Compare Description Drawings, CAD & Specs Avail. Price
industrial automation pressure input moduleAS11900G-03
$1,519
pr4000b digital power supply and readoutPR4000BF5V3
$2,785
pr4000b digital power supply and readoutPR4000BF5V2
$2,785
pr4000b digital power supply and readoutPR4000BF2V2
$2,982
pr4000b digital power supply and readoutPR4000BS2V3
$2,729
pr4000b digital power supply and readoutPR4000BS2V2
$2,706
pr4000b digital power supply and readoutPR4000BF2V3
$3,033
pdr2000a two-channel digital power supply and readoutPDR2000AO
$999