Configuration Options

The following options are available for DA06A Baratron® Heated Capacitance Manometers
Ordering Code Example: DA06A01TCEL31BAA0H00

Configuration Option Option Code
DA06A Absolute Manometer DA06A
Ranges
1 Torr 01
Units of Measurement
(corresponds to digital output units)
Torr T
mbar M
Fittings
1/2 in. OD tube BA
8 VCR female CE
8 VCR male CF
NW16 ISO-KF GA
NW25 ISO-KF GC
Mini-CF HA
Sensor Type
Standard S
Fluorine/Deposition Friendly L
Input/Output Voltages
±15 VDC 2
±24 VDC 3
Sensor Temperature
45°C 4
80°C 8
100°C 1
Electrical Connector
1
15-pin D with Screw Locks B
15-pin D with Slide Lock Posts P
Integral Relays
2
No Relays 00
Trip Point A Above Set Point, Trip Point B Above Set Point AA
Trip Point A Above Set Point, Trip Point B Below Set Point AB
Trip Point A Below Set Point, Trip Point B Below Set Point BB
Trip Point A Below Set Point, Trip Point B Above Set Point BA
Reserved
Reserved for Future Use 0
Calibration Orientation
Horizontal H
Vertical V
Accuracy
Standard 0
Other Options
None 0

1 For CE compliance, the mating electrical connector must be properly grounded.

2 Units with trip points have default set points of 50% of Full Scale, a default hysteresis of 5%, and actuation direction based on the part number code. The settings are user adjustable through EtherCAT.

* Custom part numbers can be requested for copy exact applications.

** Standard part numbers will ship with the latest firmware at the date of manufacture. A custom part number should be requested for locked firmware/EtherCAT ESI file.

Description
DA06A 3 Decade Heated Baratron® Capacitance Manometer with EtherCAT®

DA06A

Manometer, 3-Decade Temperature-Controlled EtherCAT®

Specifications

  • Type
    Temperature-Controlled Capacitance Manometer
  • Accuracy
    1000 mT to 10mT: ±0.4% of Reading
    8 mT: ±0.5% of Reading
    5 mT: ±0.7% of Reading
    3 mT: ±0.8% of Reading
    2 mT: ±1.0% of Reading
    1 mT3: ±1.5% of Reading
  • Response Time
    Pressure: 12 ms nominal (10-90% of a step response) influenced by step size and digital pressure filter setting
    EtherCAT: Refresh Rate 1 ms
  • Fitting Type
    0.50 in. OD tube, 8 VCR Female, 8 VCR Male, NW16-KF, NW25-KF, Mini-CF (1.33 in.)
    Contact MKS if alternative fitting is required.
  • Output Signal
    EtherCAT®
  • Electrical Connector
    Power: 15-pin D-subminiature male
    EtherCAT: Two RJ45 female receptacles for incoming and network signals
  • Zero Temperature Coefficient
    0.002% Full Scale/°C
  • Span Temperature Coefficient
    0.02% Reading/°C
  • Ambient Operating Temperature
    15° to 60°C (15-40 for 45°C manometer)
  • Volume
    6.3 cm3 typical.
    Contact MKS for volumes with specific flanges.
  • Warm-up Time
    4 hours
  • Overpressure Limit
    45 psia (310 kPA)
  • Materials Exposed to Process Gases
    Inconel® and Incoloy® nickel alloys (Fittings 300 Series Stainless Steel)
  • Power Requirements
    +24 VDC ±10% @ 700 mA max
    ±15 VDC ±5% @ 700 mA max
  • Indicators
    Green RUN LED and Red ERR LED for EtherCAT and Manometer status
    Green Link/Activity Indicator on bottom right of RJ45 connectors
    Rotary hex switches supporting configured station alias and explicit device identification
  • Compliance
    CE, UKCA