385 Convectron® ATM Vacuum Transducers
Model: 385620-4-GQ-T
385 Convectron® ATM Vacuum Transducers
Model: 385620-4-GQ-T

Overview
Overview
The 385620-4-GQ-T Convectron® ATM Vacuum Transducer with Display features DeviceNet communications and a 0.25 in VCR female fitting. The Convectron® ATM module combines control electronics with a piezo-resistive diaphragm sensor and the proven Convectron sensor to provide a continuous measurement from low vacuum to atmosphere. The compact design is easy to install and operate and includes a set point relay activation at a defined differential pressure from atmosphere. The Convectron ATM module eliminates the need for two sensors in those applications that require low pressure measurement and an accurate indication of differential pressure at atmosphere, such as vacuum system loadlocks. It is available with DeviceNet™ interface. The Convectron heat loss sensor provides vacuum pressure measurement over seven decades from 1x10-4 Torr to atmosphere. The diaphragm sensor technology measures the differential pressure between the inside of the vacuum space and the outside environment and activates a set point relay when a defined differential pressure is achieved. All units provide set points that are used to indicate a low vacuum level and when the defined differential pressure has been reached. This DeviceNet version with display that features a bright, 3-digit indication of the instantaneous pressure measurement.
Technical Specs
- TypeVacuum Gauge
- Measurement Range1 x 10-4 - 1000 Torr (for N2) Absolute
- Set Point Range1 x 10-4 - 1000 Torr, Absolute
- Fitting Type0.25 in VCR female
- InterfaceDeviceNet
- Connector15-pin, high density D-sub male
- Operating Temperature0 - 40°C (32° to 104°F) ambient, non-condensing
- Storage Temperature-40 - 85°C (-40 - 185°F)
- Relays4
- Relay Contact Rating1 Amp @ 30 VDC resistive, or AC non-inductive
- Exposed Materials304 stainless steel, borosilicate glass, Kovar®, alumina, NiFe alloy, polyimide, pyrex, ceramic, silicon, epoxy, RTV, Viton®, and nickel
- Case MaterialAluminum extrusion with powder coat
- Internal Volume35 cm3 (2.14 in3)
- UnitsTorr
- Display3 digit green LED, automatic ranging
- Differential Pressure Measurement Range750 Torr below Atmosphere to 250 Torr above Atmosphere
- Differential Pressure Set Point Range750 Torr below Atmosphere to 125 Torr above Atmosphere
- Differential Pressure Accuracy±2.5 Torr +2.5% of Reading
- DeviceNet VersionASCII format
- DeviceNet Baud Rates125k, 250k, 500k (Default) switch selectable
- DeviceNet Address0-63 (63 default) switch selectable
Technical Specs
- TypeVacuum Gauge
- Measurement Range1 x 10-4 - 1000 Torr (for N2) Absolute
- Set Point Range1 x 10-4 - 1000 Torr, Absolute
- Fitting Type0.25 in VCR female
- InterfaceDeviceNet
- Connector15-pin, high density D-sub male
- Operating Temperature0 - 40°C (32° to 104°F) ambient, non-condensing
- Storage Temperature-40 - 85°C (-40 - 185°F)
- Relays4
- Relay Contact Rating1 Amp @ 30 VDC resistive, or AC non-inductive
- Exposed Materials304 stainless steel, borosilicate glass, Kovar®, alumina, NiFe alloy, polyimide, pyrex, ceramic, silicon, epoxy, RTV, Viton®, and nickel
- Case MaterialAluminum extrusion with powder coat
- Internal Volume35 cm3 (2.14 in3)
- UnitsTorr
- Display3 digit green LED, automatic ranging
- Differential Pressure Measurement Range750 Torr below Atmosphere to 250 Torr above Atmosphere
- Differential Pressure Set Point Range750 Torr below Atmosphere to 125 Torr above Atmosphere
- Differential Pressure Accuracy±2.5 Torr +2.5% of Reading
- DeviceNet VersionASCII format
- DeviceNet Baud Rates125k, 250k, 500k (Default) switch selectable
- DeviceNet Address0-63 (63 default) switch selectable