The Series 943 displays pressure from 10-10 to 10-2 Torr using a high visibility LED display. Two types of analog output are standard for remote monitoring of pressure. An internal converter provides output that allows a simple log conversion from voltage to pressure over the entire operating range. A fast responding, buffered analog output is also provided. Front panel functions allow the user to disable the high voltage, as well as set a pressure at which the cold cathode will automatically turn off. This prevents the controller from displaying incorrect readings when the pressure goes above the operating range of the sensor.
Resources
Drawings & CADs
943 Cold Cathode Vacuum Gauge Controller Dimensional Drawing(202.5 kB, PDF)
Literature
943 Cold Cathode Vacuum Gauge System(305.7 kB, PDF)
Manuals
943 Digital, Vacuum Sensor System Manual(386 kB, PDF) 943 Sensor System Pin-outs(23.2 kB, PDF)