Compare Description Drawings, CAD & Specs Availability Price
baratron heated capacitance manometer with etch sensor
Heated Capacitance Manometer w/Etch Sensor, 0.1 Torr, 100°C, 8 VCR-F, DB15 w/Thread Lock, 0.50%/Rdg Accuracy, Vertical Calibration
baratron heated capacitance manometer with etch sensor
Heated Capacitance Manometer w/Etch Sensor, 0.1 Torr, 100°C, NW16 ISO-KF, DB15 w/Thread Lock, 0.50%/Rdg Accuracy, Vertical Calibration, Status LED
$3,358.00
baratron heated capacitance manometer with etch sensor
Heated Capacitance Manometer w/Etch Sensor, 10 Torr, 100°C, 8 VCR-F, DB15 w/Thread Lock, 0.25%/Rdg Accuracy, Vertical Calibration
baratron heated capacitance manometer with etch sensor
Heated Capacitance Manometer w/Etch Sensor, 10 Torr, 45°C, NW16 ISO-KF, DB15 w/Thread Lock, 0.12%/Rdg Accuracy, Vertical Calibration, Status LED
baratron heated capacitance manometer with etch sensor
Heated Capacitance Manometer w/Etch Sensor, 0.02 Torr, 45°C, NW16 ISO-KF, DB15 w/Thread Lock, 0.25%/Rdg Accuracy, Horizontal Calibration, Status LED
baratron heated capacitance manometer with etch sensor
Heated Capacitance Manometer w/Etch Sensor, 0.05 Torr, 45°C, 8 VCR-F, DB15 w/Thread Lock, 0.50%/Rdg Accuracy, Vertical Calibration

Configuration Options

Description
baratron heated capacitance manometer with etch sensor E29H Heated Capacitance Manometer, 0.02-100 Torr, Etch Sensor, 9-pin D-Sub Connector

Specifications

  • Type
    Vacuum/pressure Transducer
  • Measurement Type
    Absolute
  • Sensor Temperature
    Controlled to 45°C or 100°C
  • Full Scale Pressure Ranges
    0.02 (45°C only), 0.05, 0.1, 0.25, 1, 2, 10, 20, 100, 1000, 2000, 5000, 10000, 15000, 20000, 25000 Torr
  • Analog Output
    0-10 VDC into >10K Ω load
  • Analog Connector
    9-pin D-sub
  • Span Temperature Coefficient
    0.02% of Reading/°C
  • Operating Temperature Range
    45°C version: 15-40°C
    100°C version: 15-50°C
  • Volume
    6.3 cc
  • Warm-up Time
    2 hours for 1-25,000 Torr ranges
    4 hours for ranges < 1 Torr
  • Overpressure Limit
    45 psia (310 kPa) or 120% Full Scale, whichever is greater
  • Exposed Materials
    Inconel®
  • Power Requirements
    ±15 VDC±5% @ 0.25 Amps (0.6 Amps for 100°C version)
  • Compliance
    CE

Features

Need help?

Contact an Applications Specialist by sending us an email