355 Micro-Ion® Bayard-Alpert Vacuum Sensors
Model: 355001-YA
355 Micro-Ion® Bayard-Alpert Vacuum Sensors
Model: 355001-YA

Overview
Overview
The 355001-YA Micro-Ion® Bayard-Alpert Hot Cathode Vacuum Gauge features a Yttria-coated Iridium filament for a 5 x 10-10 - 5 x 10-2 Torr vacuum measurement range, and a 3/4 inch tube port. Granville-Phillips® 355 Micro-Ion® Hot-cathode Bayard-Alpert Ionization Vacuum Gauges are the smallest Bayard-Alpert style gauges. They are more rugged and wider range than glass gauges. In addition, they generate less heat and provide greater burnout resistance.
Technical Specs
- TypeVacuum Gauge
- Sensor TypeBayard-Alpert Hot Cathode
- Measurement Range5 x 10-10 to 5 x 10-2 Torr
- Fitting Type0.75 in. Tube
- FilamentYttria-coated Iridium
- Exposed Materials304 stainless steel, alumina, tantalum, iridium, CuAg eutectic, Kovar®
- Bake Out Temperature200°C max.
- Internal Volume10.8 cm3 (0.66 in3) to port screen
- Sensitivity20/Torr, 15/mbar, 0.15/Pa (for N2 or Air)
- X-ray Limit<3 x 10-10 Torr, <4 x 10-10 mbar, <4 x 10-8 Pa
Technical Specs
- TypeVacuum Gauge
- Sensor TypeBayard-Alpert Hot Cathode
- Measurement Range5 x 10-10 to 5 x 10-2 Torr
- Fitting Type0.75 in. Tube
- FilamentYttria-coated Iridium
- Exposed Materials304 stainless steel, alumina, tantalum, iridium, CuAg eutectic, Kovar®
- Bake Out Temperature200°C max.
- Internal Volume10.8 cm3 (0.66 in3) to port screen
- Sensitivity20/Torr, 15/mbar, 0.15/Pa (for N2 or Air)
- X-ray Limit<3 x 10-10 Torr, <4 x 10-10 mbar, <4 x 10-8 Pa