The 722C 1-25,000 Torr, General Purpose Absolute Baratron® Capacitance Manometer is a compact, low-cost manometer for accurate, repeatable, and reliable measurement. Robust sensor construction allows high overpressure tolerances, reducing errors caused by line pressure spikes. The flexible, space-saving design of the 722C makes it the ideal solution for retrofit applications as well as new equipment designs.
- Small footprint reduces system size
- Overpressure limit assures no degradation in zero repeatability or performance
- All-metal, all-welded construction exposing only Inconel® and stainless steel for high corrosion resistance
- "Drop-in" replacement for several earlier Baratron® manometers See All Features
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | 722C01TBA2FAVacuum Pressure Transducer, Absolute, 1 Torr, 0-10 VDC, 0.5 in. Tube, DB9 | |||||
![]() | 722C01TBA2FJVacuum Pressure Transducer, Absolute, 1 Torr, 0-10 VDC, 0.5 in. Tube, Terminal | |||||
![]() | 722C01TCE2FAVacuum Pressure Transducer, Absolute, 1 Torr, 0-10 VDC, 8 VCR-F, DB9 | |||||
![]() | 722C01TCE2FJVacuum Pressure Transducer, Absolute, 1 Torr, 0-10 VDC, 8 VCR-F, Terminal | |||||
![]() | 722C01TGA2FAVacuum Pressure Transducer, Absolute, 1 Torr, 0-10 VDC, NW16 ISO-KF, DB9 | |||||
![]() | 722C01TGA2FJVacuum Pressure Transducer, Absolute, 1 Torr, 0-10 VDC, NW16 ISO-KF, Terminal | |||||
![]() | 722C02TBA2FAVacuum Pressure Transducer, Absolute, 2 Torr, 0-10 VDC, 0.5 in. Tube, DB9 | |||||
![]() | 722C02TBA2FJVacuum Pressure Transducer, Absolute, 2 Torr, 0-10 VDC, 0.5 in. Tube, Terminal | |||||
![]() | 722C02TCE2FAVacuum Pressure Transducer, Absolute, 2 Torr, 0-10 VDC, 8 VCR-F, DB9 | |||||
![]() | 722C02TCE2FJVacuum Pressure Transducer, Absolute, 2 Torr, 0-10 VDC, 8 VCR-F, Terminal |
Configuration Options
The following options are available for the 722C Baratron® Absolute Capacitance Manometer.
Ordering Code Example: 722C12TCE2FJ
| Configuration Option | Option Code |
|---|---|
| 722C Absolute Capacitance Manometer | 722C |
Range Full Scale |
|
| 1 Torr | 01T |
| 2 Torr | 02T |
| 10 Torr | 11T |
| 20 Torr | 21T |
| 30 Torr | 31T |
| 50 Torr | 51T |
| 100 Torr | 12T |
| 250 Torr | RDT |
| 500 Torr | 52T |
| 1000 Torr | 13T |
| 2000 Torr (must be ordered with a fitting) | 23T |
| 5000 Torr (must be ordered with a fitting other than NW16KF) | 53T |
| 10,000 (must be ordered with a fitting other than NW16KF) | 14T |
| 20000 Torr (must be ordered with a fitting other than NW16KF) | 24T |
| 25000 Torr (must be ordered with a fitting other than NW16KF) | RCT |
Fittings (compatible with) |
|
| 0.5 in. (12.7 mm) OD tube | BA |
| 0.25 in. VCR female | CD |
| 8 VCR female | CE |
| Mini-CF, rotable | HA |
| NW16 ISO-KF | GA |
Analog Output |
|
| 0-10 VDC | 2 |
| 0-5 VDC | 3 |
Accuracy |
|
| 0.5% of Reading | F |
Electrical Connector |
|
| 9-pin Type "D" | A |
| 15-pin high density Type "D" | C |
| 4-pin bayonet type, Bendix® -compatible | D |
| 5-pin terminal strip | J |
| 15-pin Type "D" on 6-inch cable | K |
Specifications
- TypeVacuum/pressure Transducer
- Measurement TypeAbsolute
- Full Scale Pressure Ranges1-25,000 Torr (mmHg) (0.02-500 psia)
- Output0-5 VDC or 0-10 VDC into >10 kΩ load
- Accuracy0.5% of Reading
- Zero Temperature Coefficient0.008% of Full Scale/°C (10 Torr through 25,000 Torr)
0.020% of Full Scale (1 and 2 Torr)
- Span Temperature Coefficient0.04% of Reading/°C
- Operating Temperature0-50°C
15-40°C (1 and 2 Torr only) - Overpressure Limit45 psia or 2x Full Scale, whichever is greater
- Burst Pressure10x Full Scale or 100 psi, whichever is greater
- Exposed MaterialsInconel®
- ComplianceFully CE Compliant to EMC Directive 2004/108/EC1
Fully compliant to RoHS Directive 2002-95-EC
Features
Baratron® Capacitance Manometer Technology
Capacitance manometers are electro-mechanical gauges that can measure both pressure and vacuum. The capacitance gauge translates a pressure-modulated movement in a thin diaphragm into an electrical signal proportional to the pressure. The pressure sensor is the thin diaphragm that is exposed to the pressure or vacuum being measured via the inlet tube. An electrode is mounted in the reference cavity behind the diaphragm. Pressure differences between the process and the reference cavity deflect the diaphragm slightly, changing the distance between it and the electrode. Variations in this distance produce variations in the capacitance between the diaphragm and the electrode creating an electrical signal that is proportional to the pressure change. Since differences in the capacitance signal are produced by physical changes within the manometer and not by changes in the gas properties, pressure measurements by the capacitance manometer are independent of the composition of the gas being measured.
Absolute Pressure Measurement
These Baratron® pressure transducers are referenced to vacuum for absolute pressure measurement. Applications include: vacuum furnaces, freeze-drying of fruits and vegetables, gas lasers, automotive component testing, bottle coatings, and vacuum distillation.
Corrosion Resistant Inconel® Construction
All-metal, all-welded construction exposes only Inconel and stainless steel to the gas media. This allows for high overpressure tolerances that reduce errors due to line pressure spikes. High burst pressure ratings contribute to overall system safety.
Accessories
Compatible Cables
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | RCB473S-1-10Manometer Cable, PDR-5B, PDR-D-1 or PDR-D-2 Display to Manometer Terminal Block, 10 ft., RoHS, Shielded | |||||
![]() | RCB2000S-1-M3Manometer Cable, PDR2000 controller to 626, 627, 628 or 722 Manometers, DB15, 30 ft., RoHS | |||||
![]() | RCB700S-2-10Manometer Cable, PDR-5B, PDR-D-1 or PDR-D-2 Display to 722 Manometer, DB9, 10 ft., RoHS, Shielded | |||||
![]() | RCB2000S-1-M1Manometer Cable, PDR2000 controller to 626, 627, 628 or 722 Manometers, DB15, 10 ft., RoHS, Shielded | |||||
![]() | RCB112S-2-10Manometer Cable, 651, 655, 660 or PR4000 Controllers to Manometers with Terminal Strip, 10 ft., RoHS, Shielded | |||||
![]() | RCB2000S-2-M1Manometer Cable, PDR2000 Controller to 722, 750, 850 or 852 Manometers, DB9, 10 ft., RoHS | |||||
![]() | RCB2000S-2-M2Manometer Cable, PDR2000 Controller to 722, 750, 850 or 852 Manometers, DB9, 20 ft., RoHS | |||||
![]() | RCB2000S-1-M2Manometer Cable, PDR2000 controller to 626, 627, 628 or 722 Manometers, DB15, 20 ft, RoHS. | |||||
![]() | RCB2000S-2-M3Manometer Cable, PDR2000 Controller to 722, 750, 850 or 852 Manometers, DB9, 30 ft., RoHS | |||||
![]() | RCB700S-1-10Manometer Cable, 660 or PR4000 Controllers to 727, 728, 740, 750 Manometers, 10 ft., RoHS, Shielded | |||||
![]() | RCB259S-5-10Manometer Cable, 651, 655, PR4000 Controllers to 631, 624, 625, 626, 627, 628, 722 (15 pin) Manometers, DB15, 10 ft., RoHS, Shielded | |||||
![]() | 100016951Manometer Cable, 937B or 946 Controllers to 722 Manometers, DB9 Male to DB9 Female, 10 ft. | |||||
![]() | 100016952Manometer Cable, 937B or 946 Controllers to 722 Manometers, DB9 Male to DB9 Female, 25 ft. | |||||
![]() | 100016953Manometer Cable, 937B or 946 Controllers to 722 Manometers, DB9 Male to DB9 Female, 50 ft. |
Power Supplies and Displays
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | AS11900G-03Pressure Input Module, 15 pin D-Sub Connector, 4 Channel | |||||
![]() | PR4000BF2V2Digital Power Supply, Display, 2 Channel, RS232, ±15VDC, 1.5 Amp | |||||
![]() | PR4000BF5V2Digital Power Supply, Display, 2 Channel, RS485, ±15VDC, 1.5 Amp | |||||
![]() | PR4000BF5V3Digital Power Supply, Display, 2 Channel, RS485, ±24VDC, 1.0 Amp | |||||
![]() | PR4000BF2V3Digital Power Supply, Display, 2 Channel, RS232, ±24VDC, 1.0 Amp | |||||
![]() | PR4000BS2V2Digital Power Supply, Display, 1 Channel, RS232, ±15VDC, 1.5 Amp | |||||
![]() | PR4000BS2V3Digital Power Supply, Display, 1 Channel, RS485, ±24VDC, 1.0 Amp | |||||
![]() | PDR2000AODigital Power Supply, Display, 2 Channels, RS232, ±15 VDC, 0.75 Amp |
Resources
Literature
Manuals
MKS 722C Transducer Manual(724 kB, PDF) 722 Capacitance Manometer Pin-outs(20.5 kB, PDF)
Drawings & CADs
722 Baratron® Capacitance Manometer Dimensional Drawing(68.6 kB, PDF)