727A 1-1000 Torr, Heated (45°C) General Purpose Absolute Baratron® Capacitance Manometers
Model: 727A12TGA2FA
727A 1-1000 Torr, Heated (45°C) General Purpose Absolute Baratron® Capacitance Manometers
Model: 727A12TGA2FA

Overview
Overview
The 727A12TGA2FA Heated General Purpose Absolute Baratron® Capacitance Manometer is temperature controlled to 45°C and features 100 Torr Full Scale pressure measurement with a 0-10 VDC proportional analog output via a 9-pin Type "D" electrical connector, and a size NW16 ISO-KF fitting. Its blend of small size, high operating temperatures, rugged corrosion-resistant Inconel® sensor, and performance make this heated capacitance diaphragm gauge an ideal solution for new process equipment and retrofits into existing process tools. The 727 Baratron® manometer is heated to 45°C, and uses a robust sensor design that provides high tolerance to overpressure, and its all-Inconel construction gives exceptionally high corrosion resistance to process gases and by-products found in many applications.
Technical Specs
- TypePressure Transducer
- Sensor TypeCapacitance Manometer
- Measurement TypeAbsolute
- Full Scale Range100 Torr
- Sensor Temperature45°C
- Fitting TypeNW16 ISO-KF
- Analog Output0 - 10 VDC
- Electrical Connector9-pin D-sub
- Accuracy0.50% of Reading
- Response Time<75 msec
- Zero Temperature Coefficient0.010% of FS/°C
- Span Temperature Coefficient0.040% FS/°C
- Operating Temperature Range15 - 40°C
- Overpressure Limit45 psia or 2 x Full Scale range, whichever is greater
- Burst Pressure10 x Full Scale range or 90 psi, whichever is greater
- Exposed MaterialsInconel®
- Power Requirements13 - 32 VDC
Technical Specs
- TypePressure Transducer
- Sensor TypeCapacitance Manometer
- Measurement TypeAbsolute
- Full Scale Range100 Torr
- Sensor Temperature45°C
- Fitting TypeNW16 ISO-KF
- Analog Output0 - 10 VDC
- Electrical Connector9-pin D-sub
- Accuracy0.50% of Reading
- Response Time<75 msec
- Zero Temperature Coefficient0.010% of FS/°C
- Span Temperature Coefficient0.040% FS/°C
- Operating Temperature Range15 - 40°C
- Overpressure Limit45 psia or 2 x Full Scale range, whichever is greater
- Burst Pressure10 x Full Scale range or 90 psi, whichever is greater
- Exposed MaterialsInconel®
- Power Requirements13 - 32 VDC
Resources & Downloads
Drawings & CADs
Manuals
727 & 728 Capacitance Manometers Pin-outs(20.6 kB, PDF) 727 and 728 Baratron® Capacitance Manometers Instruction Manual(551.7 kB, PDF)
Literature
727A & 728A Baratron® Capacitance Manometers Datasheet(161.2 kB, PDF)