The 220DA Baratron® is temperature controlled to 45°C for use in demanding, harsh environments. The Type 220DA includes the signal conditioner and power supply, and is enclosed in a rugged, NEMA1 approved housing. It is available in ranges from 1 to 25,000 mmHg/Torr (500 psia). Accuracy is ±0.15% of Reading.
Features
Baratron® Capacitance Manometer Technology
Capacitance manometers are electro-mechanical gauges that can measure both pressure and vacuum. The capacitance gauge translates a pressure-modulated movement in a thin diaphragm into an electrical signal proportional to the pressure. The pressure sensor is the thin diaphragm that is exposed to the pressure or vacuum being measured via the inlet tube. An electrode is mounted in the reference cavity behind the diaphragm. Pressure differences between the process and the reference cavity deflect the diaphragm slightly, changing the distance between it and the electrode. Variations in this distance produce variations in the capacitance between the diaphragm and the electrode creating an electrical signal that is proportional to the pressure change. Since differences in the capacitance signal are produced by physical changes within the manometer and not by changes in the gas properties, pressure measurements by the capacitance manometer are independent of the composition of the gas being measured.
Internally Heated to 45°C
These capacitance manometers are temperature controlled to 45°C for improved accuracy. Unheated versions are exposed to ambient temperature variations which can degrade the sensor accuracy. These devices have the sensor enclosed in a volume that is maintained at a constant temperature above ambient. This solution improves the manometer’s accuracy and repeatability and lowers instrument drift by reducing or eliminating process contamination within the manometer. Heated manometers are recommended for applications that require maximum accuracy and repeatability, operate above ambient temperature and for those processes that employ hot gases.
Absolute Pressure Measurement
These Baratron® pressure transducers are referenced to vacuum for absolute pressure measurement. Applications include: vacuum furnaces, freeze-drying of fruits and vegetables, gas lasers, automotive component testing, bottle coatings, and vacuum distillation.
All-Inconel® Wetted Surfaces
These pressure transducers feature all Inconel® construction of the pressure sensor allowing it to operate without damage in virtually any chemical environment, including exposure to halogens, deionized water and steam, and ozone.
0-10 VDC and 4-20 mA Pressure Outputs
These Baratron® capacitance manometers feature two simultaneous 0 to 10 VDC and 4 to 20 mA analog pressure outputs to interface with common process control systems. The 0-10 VDC signal allows it to be used as a pressure transducer, while the 4-20 mA signal supports use as a pressure transmitter.
NEMA 1 Rated Enclosure
All pressure sensor components and electronics are housed in a rugged, NEMA 1 rated enclosure for protection when used in harsh industrial environments.
Accessories
Compatible Cables
Power Supplies and Displays
Resources
Literature
220DA Baratron® Heated Capacitance Manometer Datasheet (202.4 kB, PDF)
Manuals
220D Baratron® Capacitance Manometer Manual(562.3 kB, PDF) 220 Capacitance Manometer Pin-outs(23.2 kB, PDF)