Better than 10% valley for peaks of equal height across mass range
Bake Out Temperature
250°C with electronics removed
Operating Temperature
200°C Faraday only with electronics 10 - 40°C, 150°C Multiplier with electronics 10 - 40°C and pressure <1e-8 mbar, 90°C Multiplier with electronics 10 - 40°C and pressure <1e-5 mbar
Maximum Operating Conditions
10 - 40°C, 80% RH (non condensing)
Electronics Weight on Flange
1.7 kg
Power
24 VDC, 2A external universal voltage supply (included)
Data Acquisition Method
Dedicated realtime acquisition processor
Data Acquisition Speed
<3 ms per point for analog scans
Data Collection Methods
Analog scanning, full mass range 8, 16 or 32 points/amu, Barchart scanning, Peak-jump collection of up to 15 peaks per scan
Settling Time
<20 ms
Control Method
CE operating system processor with web-server interface
Command Structure
Documented ASCII command protocols
Electron Energy & Emission Current
40 or 70 eV fixed (set in web configuration), 1 mA
Filament Protection
Opto-isolated input for filament protect or control with auto detection of presence of the jack plug as failsafe
Communication
10/100 Base-T Ethernet, static or automatically assigned IP addresses
Enter your email address below to reset your account password.
Password Reset
Email Verification Required
Cart Items Updated
Remove Product
Remove this product from your comparison list?
Check Order Status
Provide an order number and postal code to check the status of an order or download an invoice for an order that has shipped. Login to view your complete order history.
Sign In Required
To access this and other valuable technical resources, please sign in or register for a new online account.
Add to Cart
Residual Gas Analyzer, 1-200 amu, Multichannel Plate Dual Detector, Open Ion Source, Tungsten Filaments