Type
Residual Gas Analyzer
-
Measurement Technique
Mass Spectrometry
-
Mass Range
100, 200, or 300 amu
-
Detector
Faraday
-
Faraday
1.5e-11 Torr (2e-11 mbar)
-
Filaments
Tungsten or Thoria coated Iridium
-
Software
Built-in web applications allowing RGA control and data acquisition using a platform independent web browser without the need for dedicated installed software, EasyView, Process Eye Professional and TOOLweb RGA Windows software packages for Windows 2000, XP, or Vista
-
Maximum Analyzer Operating Pressure
7.6e-5 Torr (1e-4 mbar)
-
Ion Source Sensitivity
2e-4 A/mbar
-
Resolution
Better than 10% valley for peaks of equal height across mass range
-
Bake Out Temperature
250°C (with extender or electronics removed)
-
Operating Temperature
200°C Faraday only with electronics 10 - 40°C, 150°C Multiplier with electronics 10 - 40°C and pressure <1e-8 mbar, 90°C Multiplier with electronics 10 - 40°C and pressure <1e-5 mbar
-
Cross Beam Ion Source
Single, double, or triple, depending on configuration.
-
UHV Ion Source
Single, double, or triple, depending on configuration.
-
UHV Out-gassing Rate
<1e-9 mbar -l/s
-
PVD Ion Source
Single, double, or triple, depending on configuration.
-
Zero Blast Effect
Double & Triple Filter: <5 ppm baseline levels next to mass 2 (not applicable with 5 in. single version)
-
Maximum Operating Conditions
10 - 40°C, 80% RH (non condensing)
-
Microchannel plate
3.8e-14 Torr (5e-14 mbar)
-
Single channel EM
7.6e-15 Torr (1e-14 mbar) triple configuration only
-
Electronics Weight on Flange
1.7 kg
-
Power
24 VDC, 3A external universal voltage supply (included)
-
Data Acquisition Method
Dedicated real time acquisition processor
-
Data Acquisition Speed
<3 ms per point for analog scanning
-
Data Collection Methods
Analog scanning, full mass range 8, 16 or 32 points/amu, Barchart scanning, Peak-jump collection of up to 15 peaks per scan, Complete recipe control to link multiple acquisition into a single scan, (Process Eye Professional and TOOLweb RGA only)
-
Settling Time
<20 ms (to 1 ppm of Maximum Signal)
-
Control Method
CE operating system processor with web-server interface
-
Command Structure
Documented ASCII command protocols
-
Multiplier Protection
Automatic removal of over-range peaks from all scan modes, dedicated multiplier inhibit line for over temperature or over-pressure and x-trip filament and multiplier protect for over-pressure
-
Communication
10/100 Base-T Ethernet, static or automatically assigned IP addresses
-
Interfaces
RJ45 socket: 10/100 Base-T Ethernet, 25 way D-type female: 16 configurable TTL I/O, one dedicated multiplier over temperature inhibit input, includes power for ±15V (100 mA), +24V (100 mA) and ±3.3V (100 mA), 15 way D-type female: 4 analog in (-11 - 11 V, 22 bit), 2 analog outputs (0 - 10V, 12 bit), includes power for ±15 V (100 mA), compound output from detector pre-amplifier, 9 way D-type female dedicated vacuum gauge port: -11 - 11 V 22 bit analog input, 1 digital input (gauge OK), 1 digital output (gauge enable) and gauge power +24 V (100mA), 3.5 mm jack socket: Opto-isolated input for filament protect or control with auto detection of presence of the jack plug as failsafe, 2.5 mm jack socket: Audio output for use with speaker and wired or wireless headphones
-
Number of Stored Source Settings
6 sets including ion source parameters, alignment, resolution, detector calibrations, Stored in electronics and recalled by EasyView, Process Eye Professional or TOOLweb RGA
-
Software Controlled Tuning Parameters
Electron energy, 20 - 100 eV, Emission current, 0 - 2 mA (5 mA degas), Ion energy, 0 - 10 V, Ion extraction, 0 to -130V, Filter pole bias, 0 to -10 to +10 V
-
Mounting Flange
DN 40 CF-F
-
Compliance
Compliant to RoHS Directive 2002/95/EC (control unit)